Thermo Scientific
2-1
2
Introduction
Figure 1. BF51841 Laboratory Box Furnace
The Thermo Fisher Scientific BF51841 is a reliable, energy efficient 1200 °C
laboratory box furnace. The heating elements and low thermal mass Moldatherm®
insulation provide fast duty cycles, energy conservation, and efficient
programming. Refer to
for specifications.
Features and Benefits
• Controlled heat-up rate eliminates thermal shock to materials.
• Quick heat-up and cool-down rates.
• Left swing door (reversible)
• Safety interlock switch automatically interrupts power to heating elements
when door is opened. This feature protects heating elements and eliminates
operator's exposure to electrical shock.
• Energy efficient Moldatherm insulation suitable for high interior-exterior
temperature differential. The unit is rated for a maximum operating
temperature of 1200 °C.
• Patented LGO (light gauge overbend) heating elements transfer radiant heat
energy to the work chamber and load efficiently and economically.
• Replaceable hearth plates and shelves.
• Digital control with adjustable parameters.
• Main power ON/OFF switch and power indicator on control panel.
Содержание BF51841BC
Страница 1: ......
Страница 6: ...List of Figures ii Thermo Scientific ...
Страница 18: ...Safety Notes Standards and Directives 1 12 Thermo Scientific ...
Страница 22: ...Unpacking Unpacking 3 2 Thermo Scientific ...
Страница 28: ...Installation Shelf Installation in Box Furnaces 4 6 Thermo Scientific ...
Страница 38: ...Correction of Temperature Overshoot 6 8 Thermo Scientific ...
Страница 42: ...Excess Temperature Option B Model Excess Temperature Controller Operation 7 4 Thermo Scientific ...
Страница 48: ...Communication Option Addresses for Multiple Controllers 8 6 Thermo Scientific ...
Страница 55: ...Maintenance Door Insulation Replacement Thermo Scientific 9 7 Figure 11 Door Insulation Replacement ...
Страница 57: ...Maintenance Right Hand Door Conversion Thermo Scientific 9 9 Figure 12 Door Hinge Sheet 2 2 ...
Страница 63: ...Replacement Parts Thermo Scientific 11 3 Figure 13 Wiring Diagram ...
Страница 64: ...Replacement Parts 11 4 Thermo Scientific ...
Страница 65: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable ...
Страница 66: ...Error Codes 12 2 Thermo Scientific ...
Страница 68: ...WEEE Compliance 13 2 Thermo Scientific ...
Страница 69: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable ...
Страница 70: ...Spare Parts and Accessories 14 2 Thermo Scientific ...
Страница 72: ...Device Log 15 2 Thermo Scientific ...