Thermo Scientific AquaSensors
™
AquaChlor
™
Free Chlorine Monitor User Guide
31
≡
Chlorine Config
►
Measurement
The “Measurement” screen provides a selection of Free Chlorine or HOCL
concentration.
The “pH Source” screen provides a way to select the source of the pH value in the chlorine calculation. If a pH
DataStick is detected by the AV38, this selection becomes available. If a pH DataStick is detected by the
AV38, then “pH DataStick” can also be selected. When this is done, the measured pH value is used to
calculate the selected concentration.
Note: If a pH DataStick is used to calculate the concentration, the sensor units must
be pH. If they’re not, then the concentration calculation will be done using the value
entered through the Fixed pH Value screen.
The “Fixed pH Value” screen provides a way to set a fixed pH value that can be used to calculate free chlorine
concentration from the HOCL measurement. This single fixed entry can be used when a pH DataStick is not
present or when the process pH is tightly controlled to a known value. The default value is 7.00 pH.
MENU
ESC
ENTER
◄
►
▼
AC-powered AV38
wiring
►
pH Source
►
Fixed pH Value
◄
Exit
▲
Measurement?
(ppmFCL2)
(ppmHOCL)
pH Source?
(Fixed pH)