When the value has been edited as desired, the enter key is pressed and the calibration of the point is started.
During this time, the calibration procedure can be aborted by pressing the escape key.
After the calibration point has been stored, the user is prompted to confirm a successful calibration procedure.
When the enter key is pressed, the analog output is placed into active mode and the monitoring of sensor and
temperature values is resumed. This successfully completes the temperature calibration procedure.
456.1 NTU AB
25.0 °C DS1: 6
MENU
ESC
ENTER
◄
►
▼
▲
Temperature
Confirm cal OK
MENU
ESC
ENTER
◄
▲
Temperature
Calibrating...
MENU
ESC
ENTER
◄
►
▼
▲
Temperature?
( 25.0) °C
MENU
ESC
ENTER
◄
►
▼
▲
▼
►
If an error occurs during calibration that causes the procedure to fail, the reason for the failure will be shown.