Thermo Scientific AquaSensors
™
AquaChlor
™
Free Chlorine Monitor User Guide
15
The AquaChlor
TM
measurement system is mounted on 12 inch x 12 inch panel with four ¼ inch mounting
holes. Use ¼ inch bolts with washers to mount the panel to a wall, making sure the sample chamber is
level and plumb. It is important to note that the system must be fixed securely to the wall to insure the
chamber does not move when handled or touched.
To obtain a stable chlorine measurement, the flow through the chamber should be adjusted with the inlet
valve. When properly adjusted, the right port (pH port) in the flow chamber should fill and overflow into the
center drain, while the left port (chlorine port) should partially fill and flow through a small side orifice into
the center drain. Note that the orifice connecting the chlorine port to the drain must be kept clear and not
obstructed. If debris builds up in the orifice it can be accessed for cleaning by removing the clean-out plug
in the side of the flow chamber.
.