2
Installation
Turning On the UV/Vis Detector for the First Time
Thermo Scientific
Accela UV/Vis Detector Hardware Manual
37
Turning On the UV/Vis Detector for the First Time
To turn on the UV/Vis detector for the first time
1. Ensure that the power switch at the front of the unit is in the Off position (released or out
2. Ensure that the UV/Vis detector is properly connected to the data system computer.
3. Check the fuse size.
You can operate the UV/Vis detector at 100/115 V or 230 V. However, you must ensure
that the appropriate fuses are installed (see
“Checking the Power Setting and Fuses”
4. Plug the power cord into the power receptacle on the back panel of the detector
(see
) and connect it to the power source.
5. Turn the power on by pushing in the power switch on the front panel (see
).
IMPORTANT
Before you use the UV/Vis detector for data acquisition, install the flow
cell cover, which minimizes thermal fluctuations.