Making the Back Panel Connections
28
Accela UV/Vis Detector Hardware Manual
Thermo Scientific
Figure 35.
System interconnect cable for an Accela LC system with an Accela Autosampler
To make the contact closure connections for a system with an Accela Autosampler
1. To connect the UV/Vis detector, plug one of the three connectors labeled DETECTOR
into the left, 8-pin receptacle located on the back panel of the UV/Vis detector as shown
in
2. To connect an autosampler, plug the unlabeled connector into the left, 8-pin receptacle
on the back panel of the autosampler as shown in
3. To connect an Accela pump (all models), plug a connector labeled PUMP into the 8-pin
receptacle on the back panel of the pump.
DETECT
OR
DETECT
OR
DETECT
OR
PUMP
PUMP
A/S
M/S
To
MS detector
To
autosampler
To
detector
To
detector
To
detector
To
pump
To
pump
Note
If your LC system contains more than one detector, contact your Thermo
Fisher Scientific field service engineer for instructions on how to interconnect the run
signals for the Accela LC modules.