4
Troubleshooting
Detector-Related Problems
58
Accela UV/Vis Detector Hardware Manual
Thermo Scientific
Detector-Related Problems
lists detector-related problems along with suggestions for corrective action.
Table 6.
Detector-related problems (Sheet 1 of 3)
Symptom
Cause
Remedy
Spikes on baseline
Continuous gas bubbles in the flow cell Degas the mobile phase. Connect a backpressure
device to the flow cell (check backpressure
rating).
Immiscible solvent bubbles following
mobile phase changeover
Flush the flow cell with 2-propanol and then
with mobile phase.
Electrical interference
Check the electrical lines for good connections,
interference from broadcast radiation, or both.
Check for ground loops.
Extremely large voltage fluctuations in
line power
Remove systems that cause voltage fluctuations,
for example, ovens; isolate the detector to a
“quiet” circuit; or use an uninterruptible power
supply (UPS) that is safety certified for your
country (UL, TÜV, SEMKO, DEMKO, and
so on).
Ambient light
Install the flow cell cover. You must install the
flow cell cover before turning the power on.
Never remove it when you are acquiring data.