Installation
Checking the Power Setting and Fuses
Thermo Scientific
Accela UV/Vis Detector Hardware Manual
13
Figure 12.
Fuse holder
3. Verify that the fuse holder contains the appropriate size fuses: T2.5 A for 100/115 V, or
T1.25 A for 230 V.
4. If necessary, use a pair of needle nose pliers to remove the voltage selector (see
and reinstall it so that the appropriate voltage label is upright and readable. This voltage
must agree with the incoming line voltage.
Figure 13.
Voltage selector
5. Slide the fuse holder back into the power entry module until it snaps in.
IMPORTANT
Do not connect the power cord or turn on the detector yet. Wait until you
perform the procedure in
“Turning On the UV/Vis Detector for the First Time”
to connect the power cord.
Voltage selector
Fuse holder