To complete the cleaning, the system will automatically fill the reservoir
and recirculate the bleach solution through the reservoir and system for six
hours. It will then completely drain the solution from the reservoir and
tubing by dispensing it through the system dispenser. The reservoir will
then be automatically refilled and water which will be re-circulated
through the system and again dispensed out the system dispenser to rinse
out all remaining bleach.
Once the cleaning is complete, the screen will prompt you to install a new
cartridge pack (see Cartridge Pack Installation section of this manual), and
will then begin filling the reservoir for normal use. Never install a used
cartridge pack into a cleaned system to avoid re-introducing bacteria.
Warning
This device is to be used with potable water feeds only. Cleaning
agents must be used in compliance with instructions in this manual.
Failure to comply could result in explosion and personal injury.
Warning
Be sure to connect the tubing supplied with the cleaning block
and route securely to an atmospheric drain to prevent spillage, as the
system will use this tubing to automatically drain the bleach water from
the system. Avoid splashing cleaning solution on clothing or skin. Ensure
all piping connections are tight to avoid leakage. Ensure adequate
ventilation. Carefully follow manufacturer’s safety instructions and material
safety data sheets. This device to be used with water feeds only. Cleaning
agents must be used in compliance with instructions in this manual.
Failure to comply could result in explosion and personal injury.
Cartridge packs will come bagged with four manifold connection caps.
Each cartridge pack includes one 0.2 micron absolute final filter.
1. Open right side door. Remove the four caps from the top of the
cartridge pack.
2. Verify that each of the four posts on the cartridge pack has an o-ring,
as demonstrated in Figure 3-22.
3. Remove the sanitization block and do not discard it. You’ll use this at a
later date for sanitization purposes.
4. Lift unit manifold upward, insert cartridge pack and align the pack so
that it mates with the unit manifold.
5. Lower unit manifold until it is flush with the top of the cartridge pack.
6. Hand-tighten wing head screw securely.
7. Close door.
8. DO NOT install the 0.2 micron filter and bell assembly at this time.
Type II Water System
3-17
Thermo Scientific
Section 3
Installation
System Cleaning
Procedures (cont.)
Cartridge Pack
Installation
Содержание 7165
Страница 9: ......
Страница 51: ......
Страница 67: ......
Страница 72: ...Type II Water System 7 1 Thermo Scientific WIRING DIAGRAMS ...
Страница 73: ...7 2 Type II Water System Thermo Scientific Section 7 Wiring Diagrams ...
Страница 74: ...Type II Water System 8 1 Thermo Scientific Maintenance Log Date Description Parts Needed Part Notes ...
Страница 75: ......
Страница 76: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...