A four minute manual flush of the RO membrane can be initiated through
the RO Options in the MENU. During the four minute flush, water is
sent to drain through the drain tubing and the display will show remaining
minutes. After the timed flush has ended, the TII will return to the mode
(IDLE, RUN or STBY) it was in when you initiated the flush. To initiate a
flush:
1. Choose MENU and scroll to RO Options. Press OK.
2. Choose Manual Flush and press OK.
Use the dispense knob for manual dispensing or press the DISPENSE
switch for volumetric or timed dispensing.
This feature is accessible from any screen when the volumetric dispensing
icon is an available switch.
Note
When “UNIT UNDER COUNTER” is set to “ON,” Auto Dispense
will not be available from the main unit dispenser although the manual
dispense knob will work. Volumetric dispensing is based upon total fluid
volume entering the TII. Point of dispensing accuracy will be
compromised if you are dispensing water volumetrically while
simultaneously dispensing water manually through the remote dispenser
accessory. Once Volumetric or Timed Dispensing values are set, they will
remain in memory until changed by the user. To turn the dispense method
OFF, select “OFF” in the “Dispense Method” menu.
s
1. Place a container under the dispenser.
2. Press the volumetric dispensing icon.
The screen will show current settings for Remote Volumetric, Unit
Volumetric, Timed Dispense or OFF. Press MODE to choose setting
and use the up and down arrows to set volume or time.
3. Press DISPENSE.
Type II Water System
4-5
Thermo Scientific
Section 4
Using Your TII
Manual Flush
(RO Membrane)
Dispensing Water
Automatic Dispensing
Содержание 7165
Страница 9: ......
Страница 51: ......
Страница 67: ......
Страница 72: ...Type II Water System 7 1 Thermo Scientific WIRING DIAGRAMS ...
Страница 73: ...7 2 Type II Water System Thermo Scientific Section 7 Wiring Diagrams ...
Страница 74: ...Type II Water System 8 1 Thermo Scientific Maintenance Log Date Description Parts Needed Part Notes ...
Страница 75: ......
Страница 76: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...