1. Remove the RO membrane from its packaging. Note the location of
the FEED, PRODUCT, and REJECT connections on the membrane.
The correct orientation for the installed membrane will be vertical,
with the Feed connection on the bottom. See Figure 3-9A.
2. Open the left side door on the Thermo Scientific TII.
3. Referencing Figure 3-9A, locate the three unconnected tubes (two with
elbows, one without) labeled FEED (A), PRODUCT (B) and
REJECT (C). Refer to the Push-to-Connect Fitting Tubing Install in
this section and mark FEED tubing 3/4” (1.9 cm) from end.
4. Make FEED (A) connection by first wetting the tube with water and
pushing it firmly into the membrane connector.
5. Using the Velcro® strap, secure the membrane into the right most
position against the bracket.
6. Make the PRODUCT (B) and REJECT (C) connections by first
wetting the tube stem of the elbow with water and pushing it firmly
into the membrane connector.
7. The door can remain open until the unit has been operating and you
have checked for possible water leaks at the membrane connections.
Velcro® is a registered trademark of Velcro Industries, B.V. Ltd Liab Co.
Netherlands Castorweg 22-24
Type II Water System
3-9
Thermo Scientific
Section 3
Installation
12 LPH Units
(1 Membrane)
Figure 3-9b.
24 lph RO
Membrane Installation
Figure 3-10.
RO Membrane Housing Label
Содержание 7165
Страница 9: ......
Страница 51: ......
Страница 67: ......
Страница 72: ...Type II Water System 7 1 Thermo Scientific WIRING DIAGRAMS ...
Страница 73: ...7 2 Type II Water System Thermo Scientific Section 7 Wiring Diagrams ...
Страница 74: ...Type II Water System 8 1 Thermo Scientific Maintenance Log Date Description Parts Needed Part Notes ...
Страница 75: ......
Страница 76: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...