Prior to beginning the cleaning procedure, ensure the Thermo Scientific
TII is attached to a suitable reservoir and that all water and electrical
connections between the Thermo Scientific TII and the reservoir are in
place. The system should be cleaned before the first use, before a new
cartridge pack is installed, or whenever the one-year cleaning timer alerts
you. Once set up, the cleaning procedure will run automatically. It is best
to perform the procedure when the system or water will not be required as
the entire procedure can take up to 16 hours to complete. A cleaning block
(BK1403X2) should already be installed, see “Cleaning Preparation” in this
section for details. You will also need a quantity of household bleach per
the chart on following page.
Do not use bleach with added fragrances. Cleaning your Thermo Scientific
TII system as follows:
1. Verify “Cleaning Preparation” section has been completed to prepare
your system for cleaning.
2. Press the “Menu” key, and use the up and down arrow keys to select
“Sanitization”.
3. Follow the prompts on the screen. These will guide you through the
process. It is very important that the steps are followed in order for the
cleaning to be successful. Be sure to connect the tubing supplied with
the cleaning block and route securely to an atmospheric drain to
prevent spillage as the system will use this tubing to automatically drain
the bleach water from the system. It is also important that any
equipment connected to the optional distribution loop be disconnected
from the loop and the lines plugged to prevent bleach from entering
those systems or leaking from the distribution loop.
4. When prompted to add bleach, add per the following chart:
Once the bleach has been added, if the cleaning cycle is cancelled the
system will not be available until it thoroughly rinses all of the bleach from
the reservoir and tubing. This will take up to 8 hours.
3-16
Type II Water System
Thermo Scientific
Section 3
Installation
System Cleaning
Procedures (cont.)
Bleach
Concentration
Amount needed for
30L reservoir
Amount needed for
60L reservoir
3%
240 mL (1 cup)
480 mL (2 cups)
5.25% - 6%
120 mL (1/2 cup)
240 mL (1 cup)
Содержание 7165
Страница 9: ......
Страница 51: ......
Страница 67: ......
Страница 72: ...Type II Water System 7 1 Thermo Scientific WIRING DIAGRAMS ...
Страница 73: ...7 2 Type II Water System Thermo Scientific Section 7 Wiring Diagrams ...
Страница 74: ...Type II Water System 8 1 Thermo Scientific Maintenance Log Date Description Parts Needed Part Notes ...
Страница 75: ......
Страница 76: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...