7. Follow the prompts on the screen to rinse membranes. The prompts
will ensure that you do not have a tank connected to the TII unit by
asking if the tank is empty and if the tubing from the tank goes to
drain. Answer yes to both questions. Once initiated, the rinse
procedure will last two hours, and will stop automatically.
Caution
Do not allow the Thermo Scientific TII to operate unless water is
available to the unit.
s
Warning
Use a properly grounded electrical outlet of correct voltage and
current handling capacity.
s
After performing an RO membrane rinse, connect the TII to a storage
reservoir. Refer to the Thermo Scientific TII Storage Reservoir manual.
Note
Install reservoir at the same height or higher than the TII system.
Outlet of the reservoir must be at or above the inlet of the TII.
s
3-14
Type II Water System
Thermo Scientific
Section 3
Installation
RO Membrane Rinse
(continued)
Reservoir Connection
Figure 3-19.
Thermo Scientific TII and Reservoir System Set-Up