Optional Equipment
Connecting External Devices
9-8
49iQ Instruction Manual
Thermo Scientific
4.
There are six columns for each of the six available output channels:
●
Output (V):
Displays the actual output level at the terminal of the
analog output board. For analog voltage, this value will default at zero.
For analog current, this value will default at 4 mA.
●
Decrease
,
Decrease
,
and Decrease
:
Decreases the output by
coarse, medium, or fine amounts.
●
Increase
, Increase
, and
Increase
:
Increases the output by
coarse, medium, or fine amounts.
●
Commit:
Accepts the changes to the analog output levels.
5.
For the desired analog output channel, increase or decrease the output
until the reading on the recording device indicates the proper value.
6.
After making changes to the output levels, the commit button will turn
green. To accept the changes, press the Commit button. To revert to
the previous values, press the back button to return to the previous
analog output calibration screen.
Содержание 49iQ
Страница 1: ...49iQ Instruction Manual Ozone Analyzer 117433 00 1Sep2019 ...
Страница 2: ......
Страница 8: ......
Страница 20: ......
Страница 36: ...Operation Instrument Display 3 8 49iQ Instruction Manual Thermo Scientific ...
Страница 168: ......
Страница 172: ...Maintenance Pump Rebuilding 5 4 49iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 190: ...Servicing Fan Replacement 7 8 49iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 49iQ Instruction Manual 7 49 Figure 7 56 Ozone DMC install pt 2 ...
Страница 242: ......
Страница 260: ......
Страница 268: ......
Страница 280: ......
Страница 281: ......