Troubleshooting
Troubleshooting Guide
6-4
49iQ Instruction Manual
Thermo Scientific
Problem
Possible Cause
Action
Lamp
Remove one cell and look for blue
light in hole of input block.
Lamp heater
Check lamp temperature from the
Status & Alarms>Photometer Bench
screen.
+/- 15V power supply
Check +/- 15V from the Status &
Alarms>Photometer Bench screen.
Cell A or B noise
excessive
Foreign material in one cell
Clean cell.
Defective detector
Interchange detectors at photometer
DMC PCB connectors to determine if
detector is defective.
Cell A and B noise
excessive
Foreign material in cells
Clean cells.
Lamp
failure
+/- 15V power supply
Check +/- 15V from the Status &
Alarms>Photometer Bench screen.
Output signal noisy
Recorder
Replace or repair recorder.
Sample is varying
Run instrument on stable ozone
source. If quite, no malfunction.
Foreign material in cell
Clean cell.
Sticky solenoid valve
Replace with known-good solenoid
valve.
Analyzer does not
calibrate properly
Leak
Check all tubing connections for leak.
Contaminated scrubber
Perform scrubber efficiency test.
Replace if needed.
Flow/Pressure out of
calibration
Recalibrate flow/pressure in the
operator’s manual.
Dirty system
Clean cells and flow components.
Solenoid defective
Replace with known-good solenoid
valve.
Slow response
Averaging time
Verify averaging time is set properly.
Contaminated photometer
Clean bench and then condition
system overnight.
Alarm – O
3
Lamp
temp
Defective lamp heater
Replace ozonator, lamp driver board
or lamp assembly.
Alarm –Lamp temp
Defective lamp heater
Replace bench lamp driver board or
lamp assembly.
Alarm – Flow
Flow low
Check sample capillary for blockage.
Replace as necessary.
Содержание 49iQ
Страница 1: ...49iQ Instruction Manual Ozone Analyzer 117433 00 1Sep2019 ...
Страница 2: ......
Страница 8: ......
Страница 20: ......
Страница 36: ...Operation Instrument Display 3 8 49iQ Instruction Manual Thermo Scientific ...
Страница 168: ......
Страница 172: ...Maintenance Pump Rebuilding 5 4 49iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 190: ...Servicing Fan Replacement 7 8 49iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 49iQ Instruction Manual 7 49 Figure 7 56 Ozone DMC install pt 2 ...
Страница 242: ......
Страница 260: ......
Страница 268: ......
Страница 280: ......
Страница 281: ......