Servicing
Scrubber Replacement
7-38
49iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the scrubber (
Figure 7–44
).
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Remove temporary shipping nylon tie.
3.
Twist opposite direction to unlock tube clamps.
Note
Push in tube clamp to lock.
▲
4.
Disconnect 2X plumbing.
5.
Pull scrubber out.
Figure 7–44.
Replacing the Scrubber
Scrubber
Replacement