Operation
Calibration
Thermo Scientific
49iQ Instruction Manual
3-21
The Adjust Span Coefficient screen is used to manually adjust the span
coefficient.
The following screen is shown in single range mode and dual or auto range
mode. In dual or auto range modes, “High Range” or “Low Range” is
displayed to indicate the calibration of the high or low coefficient. The
Adjust High Range Span Coefficient and Adjust Low Range Span
Coefficient screens function the same way.
Home Screen>Calibration>Advanced>Manual Calibration>Adjust
Span Coefficient (single range mode)
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust High Range Span Coefficient (dual or auto range
mode)
The Adjust Span Coefficient screen contains the following information:
●
Adjust Span Coefficient:
User manually adjusts span coefficient when in
single range mode.
●
Adjusted Concentration:
Read only. Shows adjusted concentration based
on adjusted span coefficient when in single range mode.
Adjust Span Coefficient
Содержание 49iQ
Страница 1: ...49iQ Instruction Manual Ozone Analyzer 117433 00 1Sep2019 ...
Страница 2: ......
Страница 8: ......
Страница 20: ......
Страница 36: ...Operation Instrument Display 3 8 49iQ Instruction Manual Thermo Scientific ...
Страница 168: ......
Страница 172: ...Maintenance Pump Rebuilding 5 4 49iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 190: ...Servicing Fan Replacement 7 8 49iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 49iQ Instruction Manual 7 49 Figure 7 56 Ozone DMC install pt 2 ...
Страница 242: ......
Страница 260: ......
Страница 268: ......
Страница 280: ......
Страница 281: ......