Optional Equipment
Connecting External Devices
9-6
49iQ Instruction Manual
Thermo Scientific
Table 9–4.
Analog Voltage and Current Assignment
Terminal Number
Signal Name
1
Current Out 1
2
Voltage Out 1
3 C/V
Return
1
4
Current Out 2
5
Voltage Out 2
6 C/V
Return
2
7
Current Out 3
8
Voltage Out 3
9 C/V
Return
3
10
Current Out 4
11
Voltage Out 4
12 C/V
Return
4
13
Current Out 5
14
Voltage Out 5
15 C/V
Return
5
16
Current Out 6
17
Voltage Out 6
18 C/V
Return
6
The iQ series instruments provide for the ability to calibrate the analog
outputs (both voltage and current) of the instruments. The basic procedure
for both voltage and current are the same using the following procedure:
●
Complete the connections of the recording device to the desired analog
output channel. (See page
9-5
for the channel information).
●
Calibrate the output channel low level.
Note
When calibrating the current output when using the 0-20 mA
scale, the low level will be set to 4 mA due to the inability to adjust the
actual current output to below zero.
▲
●
Calibrate the output channel full scale.
Analog Output
Calibration
Содержание 49iQ
Страница 1: ...49iQ Instruction Manual Ozone Analyzer 117433 00 1Sep2019 ...
Страница 2: ......
Страница 8: ......
Страница 20: ......
Страница 36: ...Operation Instrument Display 3 8 49iQ Instruction Manual Thermo Scientific ...
Страница 168: ......
Страница 172: ...Maintenance Pump Rebuilding 5 4 49iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 190: ...Servicing Fan Replacement 7 8 49iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 231: ...Servicing Optional DMC Ozonator Thermo Scientific 49iQ Instruction Manual 7 49 Figure 7 56 Ozone DMC install pt 2 ...
Страница 242: ......
Страница 260: ......
Страница 268: ......
Страница 280: ......
Страница 281: ......