Maintenance
Optical Bench Cleaning
5-6
49iQ Instruction Manual
Thermo Scientific
Use the following procedure to clean the mirrors.
Equipment required:
Hex drive, 7/64
1.
Unfasten nuts with hands. No tool required.
Figure 5–3.
Absorption Cleaning Removing Nuts
2.
Pull up then pull out the absorption tube.
Figure 5–4.
Absorption Tube Removal
Optical Bench
Cleaning