29 cu ft Environmental Chamber 1-3
Thermo Scientific
Section 1
Installation and Start-Up
Install the Unit
The Model 3960 Series Reach-In Incubator has four basic modes that
allow incubator setup: Run, Setpoints, Calibration and System
Configuration.
•
Run is the default mode during normal operation.
•
Set is used to enter system setpoints.
•
Calibration is used to calibrate various system parameters.
•
Configuration allows for custom setup of various options.
The chart below shows the selections under each of the modes.
Scroll for Parameters Arrows
: Steps the operator through the parameters
of SET, CAL and CONFIG Modes. The right arrow goes to the next
parameter, the left arrow returns to the previous parameter.
Up Arrow
: Increases or toggles the parameter values that have been select-
ed in the SET, CAL, and CONFIG Modes.
Enter
: Must press Enter key to save to memory all changed values.
Down Arrow
: Decreases or toggles the parameter values that have been
selected in the SET, CAL and CONFIG Modes.
Silence Key
: Press to silence the audible alarm. See Section 4 for alarm
ringback times.
Units must be installed against a wall or similar structure. Maintain a three
inch clearance behind the incubator for electrical connections.
Locate the unit on a firm level surface capable of supporting the unit’s
weight of approximately 500 lbs. Position it away from doors and windows
and heating and air conditioning ducts.
Warning
This incubator weighs approximately 500 lbs. Have sufficient
personnel available when moving.
s
Keypad Operation
RUN
SETPOINT
CALIBRATION
CONFIGURATION
Default Mode
Temperature
Temp Offset
Audible
Overtemp
Access Code
Temp Lo Alarm
Temp Relay
RS485 Address
Содержание 3960
Страница 14: ...29 cu ft Environmental Chamber 1 7 Thermo Scientific Section 1 Installation and Start Up ...
Страница 15: ......
Страница 18: ...29 cu ft Environmental Chamber 2 3 Thermo Scientific Section 2 Calibration ...
Страница 19: ......
Страница 23: ...3 4 29 cu ft Environmental Chamber Thermo Scientific Section 3 Configuration ...
Страница 29: ......
Страница 37: ......
Страница 39: ...8 2 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Страница 42: ...29 cu ft Environmental Chamber 8 5 Thermo Scientific Section 8 Parts ...
Страница 43: ...8 6 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Страница 44: ...29 cu ft Environmental Chamber 8 7 Thermo Scientific Section 8 Parts ...
Страница 45: ......
Страница 46: ...29 cu ft Environmental Chamber 9 1 Thermo Scientific Section 9 Electrical Schematics ...
Страница 47: ...9 2 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Страница 48: ...29 cu ft Environmental Chamber 9 3 Thermo Scientific Section 9 Electrical Schematics ...
Страница 49: ...9 4 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Страница 50: ...29 cu ft Environmental Chamber 10 1 Thermo Scientific Section 10 Appendix ...
Страница 51: ......
Страница 54: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...