29 cu ft Environmental Chamber 7-3
Thermo Scientific
Section 7
Specifications
Safety Specifications
. . . . . . . . . . . . . . . . . . . . . . . . .
Altitude . . . . . . . . . . . . . . . . . . . . . . .2,000 meters
Temperature . . . . . . . . . . . . . . . . . . . .5°C to 40°C
Humidity
80% RH at or below 31°C, decreasing linearly to 50% RH at 40°C
Mains Supply Fluctuations
Operating Voltage Range
Installation Category II
Pollution Degree
2
Class of Equipment
I
1 Installation category (overvoltage category) defines the level of transient overvoltage
which the instrument is designed to withstand safely. It depends on the nature of the
electricity supply and its overvoltage protection means. For example, in CAT II
which is the category used for instruments in installations supplied from a supply
comparable to public mains such as hospital and research laboratories and most
industrial laboratories, the expected transient overvoltage is 2500V for a 230V sup-
ply and 1500V for a 120V supply.
2 Pollution degree describes the amount of conductive pollution present in the oper-
ating environment. Pollution degree 2 assumes that normally only non-conductive
pollution such as dust occurs with the exception of occasional conductivity caused by
condensation.
Содержание 3960
Страница 14: ...29 cu ft Environmental Chamber 1 7 Thermo Scientific Section 1 Installation and Start Up ...
Страница 15: ......
Страница 18: ...29 cu ft Environmental Chamber 2 3 Thermo Scientific Section 2 Calibration ...
Страница 19: ......
Страница 23: ...3 4 29 cu ft Environmental Chamber Thermo Scientific Section 3 Configuration ...
Страница 29: ......
Страница 37: ......
Страница 39: ...8 2 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Страница 42: ...29 cu ft Environmental Chamber 8 5 Thermo Scientific Section 8 Parts ...
Страница 43: ...8 6 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Страница 44: ...29 cu ft Environmental Chamber 8 7 Thermo Scientific Section 8 Parts ...
Страница 45: ......
Страница 46: ...29 cu ft Environmental Chamber 9 1 Thermo Scientific Section 9 Electrical Schematics ...
Страница 47: ...9 2 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Страница 48: ...29 cu ft Environmental Chamber 9 3 Thermo Scientific Section 9 Electrical Schematics ...
Страница 49: ...9 4 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Страница 50: ...29 cu ft Environmental Chamber 10 1 Thermo Scientific Section 10 Appendix ...
Страница 51: ......
Страница 54: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...