29 cu ft Environmental Chamber 6-3
Thermo Scientific
Section 6
Factory Installed Options
To wire in the analog output board, a shielded 22 gauge, 3-conductor wire,
Part # 73041, is recommended. This wire is also readily available from
other vendors including Alpha, Part #2403, and Deerborn, Part # 972203.
Caution
Accuracy of the output at the board terminal strip to the
incubator display is ±1 unit. There is no calibration from the incubator.
Calibration to the incubator display must be at the instrument connected
to the output board.
s
To access the analog board:
1. Turn off the incubator and unplug it from the power outlet.
2. Remove the top of the incubator.
3. Locate the Analog Output board.
4. Strip the ends of the conductor and wire it to the appropriate terminals
of connectors J2 on the analog board. Refer to Figure 6-4.
Connect the Analog
Output Boards (cont.)
Figure 6-3.
Location on Back
TEMP
CO2
+
+
_
_
_
_
+
+
RH
O2
COM
J 1
J 3
J 2
TEMP
CO
2
Figure 6-4.
Connectors on Board
Содержание 3960
Страница 14: ...29 cu ft Environmental Chamber 1 7 Thermo Scientific Section 1 Installation and Start Up ...
Страница 15: ......
Страница 18: ...29 cu ft Environmental Chamber 2 3 Thermo Scientific Section 2 Calibration ...
Страница 19: ......
Страница 23: ...3 4 29 cu ft Environmental Chamber Thermo Scientific Section 3 Configuration ...
Страница 29: ......
Страница 37: ......
Страница 39: ...8 2 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Страница 42: ...29 cu ft Environmental Chamber 8 5 Thermo Scientific Section 8 Parts ...
Страница 43: ...8 6 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Страница 44: ...29 cu ft Environmental Chamber 8 7 Thermo Scientific Section 8 Parts ...
Страница 45: ......
Страница 46: ...29 cu ft Environmental Chamber 9 1 Thermo Scientific Section 9 Electrical Schematics ...
Страница 47: ...9 2 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Страница 48: ...29 cu ft Environmental Chamber 9 3 Thermo Scientific Section 9 Electrical Schematics ...
Страница 49: ...9 4 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Страница 50: ...29 cu ft Environmental Chamber 10 1 Thermo Scientific Section 10 Appendix ...
Страница 51: ......
Страница 54: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...