29 cu ft Environmental Chamber 4-1
Thermo Scientific
Section 4
Alarms
The Model 3960 Series incubator is equipped with a system which notifies
the user of an alarm condition inside the incubator. All alarms are
displayed in the control panel message center. The following table contains
information on all possible systems alarms.
To avoid alarms going off in day-to-day use, some alarms are equipped
with a time-delay feature. For this to function correctly, the alarm
condition must exist for the specified length of time before the message
center will display the alarm. This allows for interruptions, such as door
openings, to occur without the incubator going into a continuous state of
alarm.
When an alarm condition exists, the Silence key can be pressed to
temporarily mute the audible alarm. The message center will continue to
show the alarm condition. If the alarm condition is not corrected within a
specified length of time, the alarm will sound again or “ringback” to
remind the user.
When multiple alarm conditions occur, active messages are displayed in the
display center one at a time, updating at 5-second intervals. Pressing
Silence during multiple alarms causes all active alarms to be muted and to
ringback in 15 minutes.
The temperature alarms are disabled when the Temp set point is 10°C.
Table 4-1.
Systems Alarms
Description
Message Code
Delay
Ringback
Relay
No alarm condition exists
SYSTEM OK
-----
-----
-----
Temp > Otemp Set point
SYS IN OTEMP
0 min.
15 min.
Yes
Air Temp Sensor Fault
AIR SNSR ERR
0 min.
15 min.
No
Temperature Controller Failure
TMP CTRL ERR
0 min.
15 min.
Yes
Door is Open
DOOR IS OPEN
15 min.
15 min.
Yes
Temp < Temp Low Tracking Alarm
TEMP IS LOW
15 min.
15 min.
Programmable
Содержание 3960
Страница 14: ...29 cu ft Environmental Chamber 1 7 Thermo Scientific Section 1 Installation and Start Up ...
Страница 15: ......
Страница 18: ...29 cu ft Environmental Chamber 2 3 Thermo Scientific Section 2 Calibration ...
Страница 19: ......
Страница 23: ...3 4 29 cu ft Environmental Chamber Thermo Scientific Section 3 Configuration ...
Страница 29: ......
Страница 37: ......
Страница 39: ...8 2 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Страница 42: ...29 cu ft Environmental Chamber 8 5 Thermo Scientific Section 8 Parts ...
Страница 43: ...8 6 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Страница 44: ...29 cu ft Environmental Chamber 8 7 Thermo Scientific Section 8 Parts ...
Страница 45: ......
Страница 46: ...29 cu ft Environmental Chamber 9 1 Thermo Scientific Section 9 Electrical Schematics ...
Страница 47: ...9 2 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Страница 48: ...29 cu ft Environmental Chamber 9 3 Thermo Scientific Section 9 Electrical Schematics ...
Страница 49: ...9 4 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Страница 50: ...29 cu ft Environmental Chamber 10 1 Thermo Scientific Section 10 Appendix ...
Страница 51: ......
Страница 54: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...