5. Route the wires through the auxiliary hole located on the back of the
unit. See Figure 6-3 for the location of this hole and Figure 6-5 for
routing details.
6. When wiring is completed, replace the unit top and return the unit to
service.
Align the shelves and shelf channels with each of the inner doors to
facilitate introduction and removal of trays. Clean these doors with the
same care as the single door (Section 5).
Shaker support shelves are reinforced and secured to the walls of the
incubator. They have a load limit of (1) shaker or 200 pounds per shelf,
one shelf being the floor of the unit. The shaker platform limit is 50
pounds. Shakers must not exceed 250 rpm when used inside this
incubator. For shaker power connection, an internal outlet in the upper
right corner of the back wall is installed. Casters are installed at the factory
for moving the cabinet to the desired location. After the unit is in place
and prior to operation, the casters must be removed. The large rubber
vibration feet, factory installed, are positioned correctly for operation. Do
not adjust.
Caution
Any equipment placed inside the chamber must be rated for unit
operating temperature.
s
6-4
29 cu ft Environmental Chamber
Thermo Scientific
Section 6
Factory Installed Options
Connect the Analog
Output Boards (cont.)
Forma
Fuse 1 amp ( .5 amp 3961)
Fuse 2.5 amp (1.5 amp 3961)
Analog Board
RS 485 Connection
Alarm Contacts
Blower Motor
CO Sensor
2
Micro Board
Power Supply
Figure 6-5.
Board Location in Drawer
Inner Doors
Shaker Support
Shelves
Содержание 3960
Страница 14: ...29 cu ft Environmental Chamber 1 7 Thermo Scientific Section 1 Installation and Start Up ...
Страница 15: ......
Страница 18: ...29 cu ft Environmental Chamber 2 3 Thermo Scientific Section 2 Calibration ...
Страница 19: ......
Страница 23: ...3 4 29 cu ft Environmental Chamber Thermo Scientific Section 3 Configuration ...
Страница 29: ......
Страница 37: ......
Страница 39: ...8 2 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Страница 42: ...29 cu ft Environmental Chamber 8 5 Thermo Scientific Section 8 Parts ...
Страница 43: ...8 6 29 cu ft Environmental Chamber Thermo Scientific Section 8 Parts CCS ...
Страница 44: ...29 cu ft Environmental Chamber 8 7 Thermo Scientific Section 8 Parts ...
Страница 45: ......
Страница 46: ...29 cu ft Environmental Chamber 9 1 Thermo Scientific Section 9 Electrical Schematics ...
Страница 47: ...9 2 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Страница 48: ...29 cu ft Environmental Chamber 9 3 Thermo Scientific Section 9 Electrical Schematics ...
Страница 49: ...9 4 29 cu ft Environmental Chamber Thermo Scientific Section 9 Electrical Schematics ...
Страница 50: ...29 cu ft Environmental Chamber 10 1 Thermo Scientific Section 10 Appendix ...
Страница 51: ......
Страница 54: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...