![Thermo Scientific 082540 Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/thermo-scientific/082540/082540_product-manual_1105345074.webp)
Appendix A
– Current Settings
Thermo Scientific
Product Manual for the ERS 500 Suppressor
Page 74 of 84
031956-11
For Research Use Only. Not for Use in Diagnostic Procedures.
Table 14
Recommended Current Settings for the Dionex AERS 500 and
Dionex AERS 500e in the AutoSuppression Recycle and External Water
Modes with legacy discrete power supplies
Column
Eluent
Eluent Flow Rate
(mL/min)
(2 mm)
(4 mm)
Recommended Current
(mA)(2 mm)
(4 mm)
AS4A-SC
1.8 mM CO
3
2-
/ 1.7 mM
HCO
3
-
0.5
2.0
50
50
AS9-HC
9.0 mM CO
3
2-
0.25
1.0
50
50
AS12A
2.7 mM CO
3
2-
/ 0.3 mM
HCO
3
-
0.38
1.5
50
50
AS14
3.5 mM CO
3
2-
/ 1.0 mM
HCO
3
-
0.3
1.2
50
50
AS14A
8.0 mM CO
3
2-
/ 1.0 mM
HCO
3
-
0.5
(1)
1.0
50
50
AS22
4.5 mM CO
3
2-
/ 1.4 mM
HCO
3
-
0.3
1.2
50
50
AS23
4.5 mM CO
3
2-
/ 0.8 mM
HCO
3
-
0.25
1.0
50
50
AS10
80 mM OH
–
0.25
1.0
50
300
AS11
12 mM OH
–
0.25
1.0
50
50
38.3 mM OH
–
0.5
2.0
50
300
AS11-HC
30 mM OH
–
0.38
1.5
50
300
60 mM OH
–
0.38
1.5
100
300
AS15
38 mM OH
–
0.3
1.2
50
300
40 mM OH
–
0.5
(1)
50
AS16
35 mM OH
–
0.25
1.0
50
100
55 mM OH
–
0.38
1.5
100
300
AS17-C
15 mM OH
–
0.25
1.0
50
50
40 mM OH
–
0.5
2.0
50
300
AS18
23 mM OH
–
0.25
1.0
50
100
39 mM OH
–
0.25
1.0
50
100
AS19
20 mM OH
–
0.25
1.0
50
50
45 mM OH
–
0.25
1.0
50
300
AS20
35 mM OH
–
0.25
1.0
50
100
55 mM OH
–
0.25
1.0
50
300
AS21
15 mM OH
–
0.35
50
AS24
55 mM OH
–
0.30
50
AS25
36 mM OH
–
0.25
1.0
50
100
AS26
55 mM OH
–
0.30
1.2
50
300
(1)
3 mm format