4
– Example Applications
Thermo Scientific
Product Manual for the ERS 500 Suppressor
Page 55 of 84
031956-11
For Research Use Only. Not for Use in Diagnostic Procedures.
4.7
Cation Example Applications
4.7.1
Dionex IonPac CS12A Blank and Standard Analysis
Sample:
100 µL of Standard or Blank neutralized in the Dionex CERS 500
Concentrator Column:
Low Pressure Trace Cation Concentrator (i.e. Dionex TCC-XLP1, Item # 063889)
Guard Column:
Dionex IonPac CG12A Guard Column
Analytical Column:
Dionex IonPac CS12A Analytical Column
Eluent:
20 mM Methanesulfonic acid
Eluent Flow Rate:
1.0 mL/min
Electrolytic
Suppressor:
Dionex Cation Electrolytically Regenerated Suppressor, Dionex CERS 500
Recycled Eluent Mode
or Chemical
Suppressor:
Dionex Cation Chemically Regenerated Suppressor, Dionex CCRS 500
Chemical Regenerant:
100 mN TBAOH
Expected Background Conductivity:
20 mM MSA: <1 µS
Expected System Operating Back Pressure:
Without Guard: 1,400 psi (9.65 MPa)
With Guard: 1,850 psi (12.75 MPa)
Analyte
Conc..
(µg/L)
1. Lithium
5.0
2. Sodium
20.0
3. Ammonium
25.0
4. Dimethylamine
B
5. Potassium
50.0
6. Trimethylamine
B
7. Magnesium
25.0
8. Calcium
50.0
Figure 9
Dionex IonPac CS12A Blank and Standard Analysis
Figure 14A
Figure 14B