2
– Installation
Thermo Scientific
Product Manual for the ERS 500 Suppressor
Page 22 of 84
031956-11
For Research Use Only. Not for Use in Diagnostic Procedures.
The Dionex ERS 500, Dionex ERS 500e or Dionex AERS 500 Carbonate is installed in the
column compartment or detector compartment of the chromatography module immediately after
the analytical column and before the conductivity detector cell. All components required for
installation of the suppressor are included with the system.
Gas Separator Waste Tube (Item # 045460)
Backpressure coil(s)
o
4 mm (Item # 045877)
o
2 mm (Item # 045878)
Microbore Tubing (2 mm only) (Item # 052324)
Mounting clip (Item # 045612)
Options:
Dionex CRS/ERS Installation Kit (Item # 038018)
(Pressurized Water Delivery System used with AutoSuppression External Water Mode,
Chemical Suppression Mode, or MPIC Suppression Mode).
Peristaltic Pump Kit (Item #: 064508). (Water delivery system used with
AutoSuppression External Water Mode, Chemical Suppression Mode, or MPIC Mode).
The Dionex ERS 500, Dionex ERS 500e and Dionex AERS 500 Carbonate must be
operated with the Gas Separator Waste Tube (Item # 045460)
The use of ¼-28 or 10-32 ferrule/bolt style liquid lines may be required for installation and
use of ERS 500. See “Dionex Liquid Line Fittings” for complete details.
2.2
Electrolytically Regenerated Suppressor Control
Always turn the pump and the Dionex
ERS Control on and off at the same time. Eluent
flow through the Dionex
ERS 500 is required for proper operation. Without current, the
membranes and screens in the Dionex
ERS 500 will become exhausted by the flowing
eluent resulting in small analyte peak areas. If this should occur, perform the procedure in
The Dionex AERS 500 Carbonate suppressor is not compatible with discrete power
supplies and should only be operated with power supplies that can deliver current in 1 mA
increments.
If an ERS Mode is not available, select the equivalent SRS Mode on the power supply to
support the
Dionex
ERS 500 suppressor. The Dionex
ERS 500 and Dionex ERS 500e are
fully compatible with SRS settings.
CAUTION
!
NOTE
!
CAUTION
!
CAUTION
!
NOTE
!