Detection Limits of Dionex ERS 500e Sample Pretreatment and Subsequent Ion
Chromatography
4.4
Recovery Data for Trace Anions in 25% Sodium Hydroxide
Anion
25% NaOH*
Spike
Expected
Found*
(ppb)
(ppb)
(ppb)
(ppb)
Chloride
155 + 3
100
255
258 + 4
Bromide
—
100
100
99 + 8
Chlorate
—
100
100
96 + 9
Nitrate
11 + 6
100
111
119 + 13
Phosphate
106 + 25
100
206
183 + 17
Sulfate
348 + 4
200
548
553 + 11
Oxalate
134 + 28
100
234
238 + 22
* n = 8
Separation Conditions are shown in
, “Dionex IonPac AS12A Isocratic Analysis of 25% Sodium Hydroxide.”
4.5
Recovery Data for Trace Cations in 24% Sulfuric Acid
Cation
24% H
2
SO
4
*
Spike
Expected
Found*
(ppb)
(ppb)
(ppb)
(ppb)
Lithium
---
2.0
2.0
1.993 + 0.008
Sodium
7.4 + 0.4
8.0
15.4
14.3 + 0.1
Potassium
3.3 + 0.5
20.0
23.3
22.5 + 0.2
Magnesium
2.0 + 0.2
10.0
12.0
12.2 + 0.2
Calcium
10.6 + 0.1
20.0
30.6
33 + 1
* n = 8
Separation Conditions are shown in
, “Dionex IonPac CS12A Isocratic Analysis of 24% Sulfuric Acid”.
Cation
ppb*
Lithium
0.03
Sodium
1.0
Potassium
2.0
Magnesium
0.6
Calcium
0.8
* Estimated values in H
2
SO
4
matrix
Anion
ppb*
Chloride
4
Bromide
20
Chlorate
30
Nitrate
20
Phosphate
50
Sulfate
30
Oxalate
50