95 Series Operator’s Guide
4-10
71-0011 — REV E
Using the default zero value, the readings are approximate. You
should recalibrate as soon as possible, using N
2
gas for zero and
fresh air for span. The reading with N
2
may be several percent off
zero until you press the ZERO/- button.
Return to Normal Operation
After calibration, return to normal operation.
1.
Close the flow control knob or the dispensing valve.
2.
Remove the test cup and close the flap that covers the
operator controls.
3.
Disassemble the test kit. Store the components in the
storage case. You can leave the hoses attached for the next
calibration. Your instrument is now ready for normal
operation.
NOTE
While in normal mode, the HS-95 or CO-95 goes into a FAULT
condition if you press the SPAN/+ button for several seconds
without introducing calibration gas to the sensor. To return to
normal operation, turn the power switch off, wait a few seconds,
then turn the power switch back on.
If the HS-95 or CO-95 “fails” a calibration (with gas introduced
to the sensor), see page 4-9.