95 Series Operator’s Guide
4-4
71-0011 — REV E
Preparing for Calibration (all models)
NOTE
With a HS-95 or CO-95, if the battery has gone dead or the
battery has been out of the instrument for 30 minutes or more,
allow 1 hour after the new battery has been installed for the
sensor to stabilize before beginning calibration procedures.
1.
Check that the flow control knob or dispensing valve is shut
off. Screw the regulator or the dispensing valve onto the test
cylinder for your detector.
2.
Securely attach one end of the tubing to the gas outlet of the
regulator or dispensing valve. For the HS-95, securely
attach the other end of the tubing directly to the test cup.
3.
For the CO-95 or OX-95, attach the other end of the tubing
from the dispensing valve to the inlet (bottom) connector of
the flowmeter. Attach a second piece of tubing between the
outlet (upper) connector of the flowmeter and the test cup.
CAUTION
The flowmeter must be vertical and right side up in order for you
to attain accurate readings.
How to Use the Control Buttons
•
The control buttons are ZERO/- and SPAN/+.
•
A button should be pressed for at least 1/2 second to be sure
that it is recognized.
•
The ZERO/- button is used to set zero. The SPAN/+ button
is used to set span. See the Calibration section for details.