95 Series Operator’s Guide
4-8
71-0011 — REV E
3.
Press and hold the ZERO/- button. A low bar is displayed to
the left of the gas reading during the automatic zero
calibration. Release the button when the PASS or FAIL
message is displayed. The instrument returns to normal
mode, with the alarm inhibited for 2 minutes.
4.
Turn off the gas and remove the test cup.
5.
Let the instrument stabilize for 90 seconds in fresh air. Press
and hold the SPAN/+ button. A high bar is displayed to the
left of the gas reading during the automatic span calibration.
Release the button when the PASS or FAIL message is
displayed. The instrument returns to normal mode, with no
additional alarm inhibition.
NOTE
The OX-95 gas alarms are inhibited for 2 minutes after a
successful (PASS) zero calibration and after an unsuccessful
(FAIL) span calibration. This allows you to finish calibration
without setting off an alarm. During the inhibit period, a rotating
pattern of horizontal bars appears to the left of the gas reading to
remind you that the alarms are shut off. The OX-95 alarms are
NOT inhibited after a successful (PASS) span calibration.
If Calibration Fails
On all models, if the zero or span calibration fails, the FAIL
message is displayed. Try the zero or span procedure a second
time. If the FAIL message appears again:
1.
Turn the instrument off.
2.
Replace the sensor.