Weak or Failed Sensor Indications
Weak sensor: When a weak sensor is detected during
calibration, the word FAULT blinks on the LCD. The instrument
is still functioning normally, but you should replace the sensor as
soon as possible.
Failed sensor: When a defective sensor or sensor circuit is
detected, the word FAULT is displayed on the LCD. In addition,
the buzzer sounds continuously and the red alarm light is on
steady. A fault condition is caused by a missing or bad sensor, a
bad connection, or a main board sensor circuit fault.
Recommended Action:
Remove the instrument from the monitoring environment. Refer
to the troubleshooting guide in Chapter 4 to determine the cause
of the alarm condition and the recommended course of action.
WARNING
A reading of 23.5% O
2
or more must be treated as a high-
oxygen alarm condition until proven otherwise.