Calibration & Maintenance
71-0011 — REV E
4-7
6.
Press and hold the SPAN/+ button. A high bar is displayed
to the left of the gas reading during the automatic span
calibration. Release the button when the PASS or FAIL
message is displayed. The instrument returns to normal
mode, with the alarm inhibited for 30 seconds.
NOTE
The HS-95 and CO-95 gas alarm is inhibited for 2 minutes after a
successful (PASS) zero calibration and after an unsuccessful
(FAIL) span calibration. The alarm is also inhibited for
30 seconds after a successful (PASS) span calibration. The alarm
is inhibited to allow you to finish calibration without setting off
the alarm. During the inhibit period, a rotating pattern of
horizontal bars appears to the left of the gas reading to remind
you that the alarm is shut off.
Calibration (OX-95)
Open the operator controls flap, and slide the power switch on
(toward the sensor). Allow a few seconds for the instrument to
stabilize.
1.
Fit the test cup into the groove around the sensor port with
the retainer arm around the back of the instrument holding
the cup in place. The gas inlet should extend to the right
across the top of the instrument.
2.
Open the N
2
cylinder dispensing valve. Set the flow rate to
1.0 SCFH. Allow 90 seconds for the reading to stabilize.
The low oxygen alarm will sound. This is normal.
Содержание CO-95