The input is blocked if a temperature setting over I/O port
already exists or the programmer is active.
14.2.3 Setting the RTA system correction value
RTA (Real Temperature Adjustment)
The display shows the actual temperature at the internal or
external control sensor with the selected resolution.
This temperature does not correspond directly to the tem-
perature in the circulator’s bath and even less to the temper-
ature in the external connected system.
The temperature difference is determined by measuring the
actual current temperature using a suitable measuring de-
vice (calibrated or gauged thermometer).
It is entered into the circulator as the correction factor RTA
and remains stored there.
A separate internal (i) and external (e) RTA value is assigned
to each of the 4 storable set temperatures.
The correction factor only refers to this one application. A
new correction value is required for any new temperature or
altered test setup.
1
Select „Set values“ and then the set temperature
whose RTA values you want to change. Activate
„Change RTA“ and select the desired RTA value (inter-
nal/ external) and change the display in the same way
as decribed in Chapter 14.2.1.
14.3 Change actual set value quickly
The function „Change actual set value“ in the main menu al-
lows to change the currently active set temperature quickly
without going to the „Set values“ submenu first.
The changed set value is stored permanently and is acti-
vated immediately.