Internal temp.
External temp.
Set value
°
C
°
C
°
C
22.68
21.20
40.00
17:00:00 17.11.00
20.00
°
C
70.00
°
C
00:08:09
R1 S1
Operating
45
HAAKE Phoenix II P1
Version 3.07
14.4.6 Running a temperature program
The currently selected program is run by the „Start ramp“
function.
In the main menu „Stop ramp“ is displayed instead of „Start“.
After that the unit heats/ cools until the start tempe/
-- the set variation is reached. After that the first segment is
started automatically.
In the main menu the end temperature and the remaining
time of the current segment as well as the number of the
ramp and the number of the segment are displayed next to
the ramp symbol. When the course of the program has been
ended, the circulator maintains the last set temperature. The
end of a segment or the end of the program can be indicated
by an acoustic signal (see 14.4.5).
14.4.7 Stopping the temperature program
You can stop a running program with the „Stop ramp“ func-
tion. The unit now maintains constantly the last set tempera-
ture of the program.
When restarting the program, the unit heats/ cools, until the
start temperature
±
the set variation is reached. Then the
program starts again with the first segment.