Serial Interface
70
HAAKE Phoenix II P1
Version 3.07
19.8 Operating status / Error message
Call up operating status:
Standard set of commands:
B<cr> or
Extended set of commands:
R_BS<cr>
After entering one of these commands, the following twelve
”state flags” are shown:
” x x x x x x x x x x x x $ ”
1 :
Alarm external Pt100
0 :
External Pt100 OK
1 :
Alarm internal Pt100
0 :
Internal Pt100 OK
1 :
0 :
Liquid level OK
:
0 :
Reserve
1 :
Alarm cooling
0 :
Cooling OK
1 :
Alarm via external connection
0 :
External connection OK
1 :
Alarm motor or pump overloading
0 :
Motor and pump OK
1 :
Alarm liquid level
0 :
Liquid level OK
1 :
Alarm overtemperature
0 :
Temperature OK
1 :
Main relay missing
0 :
Main relay present
1 :
External control
0 :
Internal control
1 :
Temperature control ON
0 :
Temperature control OFF