Serial Interface
71
HAAKE Phoenix II P1
Version 3.07
19.9 Example definition of a ramp program by PC
The comfortable programmer functions integrated in the
Phoenix units can also be controlled by PC through the
RS232C interface.
1.
Select the
number for the program
you would like to
define (max. 10 programs possible):
W
[
RN
[
<
program no.
>
1 ... 10
Space
Bar
2.
Then define the
number of program cycles
.
A program can be repeated up to 90 times (i.e. 90 program
cycles):
W
[
RZ
[
<
number of cycles
>
1 ... 90
3.
W_SA will create a new segment.
A segment number will automatically be assigned to
the programmed segments. Now you can enter the following
parameters for each
segment number
:
•
the
start
temperature of the segment
•
the
end
temperature of the segment
•
the
run time
of the segment in seconds (
≥
1)
W
[
SP
[
<
segment no.
>
[
<
start
>
[
<
end
>
[
<
run time
>
[
<
tape
>
Example:
segment no. 1
start
=
23.0
°
C
end
=
27.0
°
C
run time =
600 sec
tape
=
0.5
°
C
W
[
SP
[
1
[
23.0
[
27.0
[
600
[
0.5
All other segments can be defined accordingly. Please note
that the temperature program is continual for all defined seg-
ments. This means that the end temperature of segment i
must be the start temperature of segment i+1.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com