Appendix B RS-232 Commands
B-25
lr
xy
,
sr
xy
, er
xy
These commands report the last long or short record stored, or the current dynamic data
record. The
x
can be a 0 or 1, 0=no check sum returned, 1= check sum returned. The
y
is
the format code, which can be 1, 2, or 3 their functions are located in Table B-1 “Format
Codes”.
Send:
“lr01”
Receive:
“
10:15 12-28-02 flags 9C040000 co 7349E+0 loco 5994E+0 innt 33.2 cht
44.7 pres 758.9 smplfl 1.085 speed 100.0 biasv –115.5 intensity 199940
”
lrec
srec
lrec
xxxx yy
srec
xxxx yy
lrec
aa:bb oo-pp-qq yy
srec
aa:bb oo-pp-qq yy
xxxx
= the
number of past records
yy
=
the number of records to return (1 to 10)
aa
= hours (01 to 24, military time)
bb
= minutes (00 to 59)
oo
= month (01 to 12)
pp
= day (01 to 31)
= year (enter "00" for the year 2000)
These commands output long or short records. The output format is determined by the
set lrec
format
and
set srec format
commands. The logging period is determined by the
set lrec per
and
set srec per
commands.
In the following example, there are 740 long records currently stored in memory. When
the command
lrec 100 5
is sent, the instrument counts back 100 records from the last
record collected (record 740), and then returns 5 records: 640, 641, 642, 643, and 644.
Send:
"lrec 100 5"
Receive:
Record 640
:
“
10:15 12-28-02 flags 9C040000 co 7349E+0 loco 5994E+0 innt 33.2 cht 44.7 pres 758.9
smplfl 1.085 speed 100.0 biasv –115.5 intensity 199940
Record 641
: 10:20 12-28 flags 9C040000 co 7349E+0 loco 5994E+0 innt 33.2 cht 44.7 pres 758.9 smplfl
1.085 speed 100.0 biasv –115.5 intensity 199940
Record 642
: 10:25 12-28 flags 9C040000 co 7349E+0 loco 5994E+0 innt 33.2 cht 44.7 pres 758.9 smplfl
1.085 speed 100.0 biasv –115.5 intensity 199940
Record 643
: 10:30 12-28 flags 9C040000 co 7349E+0 loco 5994E+0 innt 33.2 cht 44.7 pres 758.9 smplfl
1.085 speed 100.0 biasv –115.5 intensity 199940
Record 644
: 10:35 12-28 flags 9C040000 co 7349E+0 loco 5994E+0 innt 33.2 cht 44.7 pres 758.9 smplfl
1.085 speed 100.0 biasv –115.5 intensity 199940
”
Содержание 48C
Страница 4: ......
Страница 12: ......
Страница 14: ......
Страница 18: ......
Страница 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Страница 112: ......
Страница 116: ......
Страница 120: ......
Страница 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Страница 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Страница 138: ......
Страница 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Страница 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Страница 158: ......
Страница 178: ...Appendix B RS 232 Commands B 20 µ ...
Страница 184: ......
Страница 186: ...C 2 ...
Страница 187: ...C 3 ...
Страница 188: ...C 4 ...
Страница 189: ...C 5 ...
Страница 190: ...C 6 ...
Страница 191: ...C 7 ...
Страница 192: ...C 8 ...
Страница 193: ...C 9 ...
Страница 194: ...C 10 ...
Страница 195: ...C 11 ...
Страница 196: ...C 12 ...
Страница 197: ...C 13 ...
Страница 198: ...C 14 ...
Страница 199: ...C 15 ...
Страница 200: ......
Страница 201: ...APPENDIX D PUMP MAINTENANCE ...
Страница 202: ......
Страница 203: ......
Страница 204: ......