Chapter 3 Operation
3-39
Sample/Reference Ratio
The Sample/Reference Ratio screen, shown below, displays the ratio of the
intensities of the light source through the sample (CO) side and reference (N
2
)
side of the correlation wheel. Normally, when zero air is being sampled, the
sample/reference ratio is between 1.14 and 1.18. A ratio outside of this range may
indicate that the correlation wheel is dirty or leaking gas.
To display the Sample/Reference Ratio screen:
From the Main Menu choose Diagnostics
From the Diagnostics menu choose Sample/Ref Ratio
To use the Sample/Reference Ratio screen:
This is a view only screen
Press
MENU
to return to the Diagnostics menu
Press
RUN
to return to the Run screen
Sample/Reference Ratio Screen
Содержание 48C
Страница 4: ......
Страница 12: ......
Страница 14: ......
Страница 18: ......
Страница 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Страница 112: ......
Страница 116: ......
Страница 120: ......
Страница 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Страница 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Страница 138: ......
Страница 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Страница 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Страница 158: ......
Страница 178: ...Appendix B RS 232 Commands B 20 µ ...
Страница 184: ......
Страница 186: ...C 2 ...
Страница 187: ...C 3 ...
Страница 188: ...C 4 ...
Страница 189: ...C 5 ...
Страница 190: ...C 6 ...
Страница 191: ...C 7 ...
Страница 192: ...C 8 ...
Страница 193: ...C 9 ...
Страница 194: ...C 10 ...
Страница 195: ...C 11 ...
Страница 196: ...C 12 ...
Страница 197: ...C 13 ...
Страница 198: ...C 14 ...
Страница 199: ...C 15 ...
Страница 200: ......
Страница 201: ...APPENDIX D PUMP MAINTENANCE ...
Страница 202: ......
Страница 203: ......
Страница 204: ......