Chapter 3 Operation
3-19
CO Span Gas
The CO span gas is set prior to calibration. The span gas is used to calibrate the CO
level.
To display the CO Span Gas screen:
From the Main Menu choose Calibration Factors
From the Calibration Factors menu choose Span Gas
To use the CO Span Gas screen:
Use the
↑
and
↓
pushbuttons to increment/decrement the gas value
Use the
←
and
→
pushbuttons to change the gas value
Press
ENTER
to accept a change
Press
MENU
to return to the Calibration Factors menu
Press
RUN
to return to the Run screen
Single and Dual/Auto range CO Span Gas Screen
SPAN GAS:
PPM 00000.250
SET TO 00000.500
78
INC/DEC
HI SPAN GAS:
PPM 00000.250
SET TO 01000.000
78
INC/DEC
Содержание 48C
Страница 4: ......
Страница 12: ......
Страница 14: ......
Страница 18: ......
Страница 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Страница 112: ......
Страница 116: ......
Страница 120: ......
Страница 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Страница 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Страница 138: ......
Страница 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Страница 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Страница 158: ......
Страница 178: ...Appendix B RS 232 Commands B 20 µ ...
Страница 184: ......
Страница 186: ...C 2 ...
Страница 187: ...C 3 ...
Страница 188: ...C 4 ...
Страница 189: ...C 5 ...
Страница 190: ...C 6 ...
Страница 191: ...C 7 ...
Страница 192: ...C 8 ...
Страница 193: ...C 9 ...
Страница 194: ...C 10 ...
Страница 195: ...C 11 ...
Страница 196: ...C 12 ...
Страница 197: ...C 13 ...
Страница 198: ...C 14 ...
Страница 199: ...C 15 ...
Страница 200: ......
Страница 201: ...APPENDIX D PUMP MAINTENANCE ...
Страница 202: ......
Страница 203: ......
Страница 204: ......