Basic Test Setup
3.2
TSW14J56EVM Setup
See the TSW14J56EVM User's Guide (
) for a more detailed explanation of the TSW14J56
setup and its features. This document assumes that the HSDC Pro software and the TSW14J56 pattern
capture and generation board are both installed and functioning properly. This information can be found at
http://www.ti.com/tool/tsw14j56evm
.
3.3
ADS54J54EVM Quick-Start Procedure
Connect J3 of a ADS54J54EVM into J4 of a TSW14J56 JESD204B capture card.
3.3.1
TSW14J56 Data Capture Card
1. Connect a 5-V power supply to connector
J11
of the TSW14J56EVM.
2. Flip switch
SW6
to the
ON
position.
3. Insert a USB cable into the USB port on the TSW14J56. Connect the other end to the PC.
3.3.2
ADS54J54EVM
1. Connect a bench 5-V power supply, or equivalent, to the connector J13 using the supplied power
supply cable. Alternatively, power may be supplied to the EVM by way of the red banana jack J8 for +5
V and the black banana jack J9 for ground.
2. Connect a USB cable to the USB port on the ADS54J54EVM and connect the other end to the PC.
3. Connect a signal source to an analog input SMA connector such as J1 (
A_INP
). For single-tone
testing, a bandpass filter should be used to achieve the best SNR and harmonic performance.
15
SLAU616A – January 2015 – Revised January 2016
ADS54J54 Evaluation Module