6 Using the product
47
6.1.3.3.
Sensor protection
Protection limits can be set to protect the sensors against overload.
Sensor protection switch-off is available for the following sensors:
CO, NO.
The sensor protection is activated when the threshold is exceeded.
• testo 330-1 LL: Switch-off.
• testo 330-2 LL: Dilution, if exceeded again: Switch-off.
To deactivate sensor protection the thresholds must be set to 0
ppm.
Calling up the function:
>
[ ]
→
Sensor Settings
→
Sensor Protection
→
[OK]
.
Setting sensor protection thresholds:
1.
Select parameter:
[Edit]
.
2.
Set parameter
→
[OK]
.
3. Save changes:
[Finished]
.
6.1.3.4.
Recalibration / adjustment
CO and NO sensors can be recalibrated and adjusted.
For recalibration / adjustment Testo recommends the use of the
calibration adapter 0554 1205.
If obviously unrealistic readings are displayed, the sensors
should be checked (calibrated) and, if required, adjusted.
Have the recalibration / adjustment carried out by a
qualified service centre approved by Testo.
Adjustments made with low gas concentrations can lead to
accuracy deviations in the upper measuring ranges.
Calling up the function:
>
[ ]
→
Sensor Settings
→
Recalibration
→
[OK]
.
Possibly:
>
Enter the password:
[Enter]
→
Enter password
→
[Next]
→
[OK]
.
- Gas zeroing (30 s).