Specifications
5
d
e
en
fr
es
it
p
t
sv
nl
??
??
Specifications
4
Products with Bluetooth
Ò
(Option)
Changes or modifications, which are not expressly approved by the responsible
official body, can lead to a withdrawal of operating permission.
Interference with data transfer can be caused by instruments which transmit on
the same ISM band, e.g. microwave ovens, DECT telephony, non-secure
software in cellphones when telephoning, transmitting/receiving text messages
etc.
The use of radio connections is not allowed in e.g. aeroplanes and hospitals.
For this reason, the following point must be checked before entering:
Deactivate Bluetooth function
M
Maaiinn m
meennuuee -- S
Seettttiinnggss -- B
Buueettooootthh -- ddeeaaccttiivvaattee B
Blluueettooootthh ((O
Offff))
Protecting the environment
Take faulty rechargeable batteries/spent batteries to the collection points
provided for them.
Send the product back to Testo at the end of its useful life. We will ensure
that it is disposed of in an environmentally friendly manner.
Specifications
Functions and use
The testo 327 is a hand-held measuring instrument for the professional flue gas
analysis of domestic and light commercial gas and oil boilers and appliances.
This includes condensing boilers and gas heaters. The unit is designed to carry
out tests shown in BS 7967, and has a timed tightness test, let by test and a
timed CO function.
These systems can be adjusted using the testo 327 and checked for
compliance with the applicable limit values.
The testo 327 is available in four versions; the scope of function varies
according to the version:
· testo 327 O
2
:
Infrared interface
· testo 327 CO:
Infrared interface
· testo 327-1 (O2, CO):
Infrared interface
· testo 327-2 (O2, CO):
Infrared / IRDA interface, memory, automatic
sensor diagnosis, option: Bluetooth (data
interface)
Warning
The testo 327 must not be used in areas at risk of explosion,
for long-term measurements or as a safety (alarm) device!
The testo 327 with the Bluetooth option may only be operated in countries in
which it is type approved (see Technical Data).
Technical data
D
Diissppllaayy vvaarriiaabblleess [[uunniittss]]
M
Meeaassuurriinngg rraannggee//rreessoolluuttiioonnA
Accccuurraaccyy//rreessppoonnssee ttiim
mee
11))
O
Oxxyyggeenn,, vviiaa iinntteerrnnaall eelleeccttrroo--cchheem
miiccaall sseennssoorr ((nnoott 332277 C
CO
O))::
O
O
22
content [%],
O
O
22
aaiirr supply [%],
Reference value O
O
22
rreeff [%]0...21% / 0.1%
±0.2% / t90 <40s
C
Caarrbboonn m
moonnooxxiiddee,, 332277--11::
C
CO
O content [ppm, mg/m³] 0...4,000ppm / 1ppm
±20ppm (0...400ppm), ±5% of reading
(401...1,000ppm),
(H
2
share <10%)
±10% of reading (1,001...4,000ppm) / t90 <60s
C
Caarrbboonn m
moonnooxxiiddee,, vviiaa iinntteerrnnaall eelleeccttrroo--cchheem
miiccaall sseennssoorr ((oonnllyy 332277--22 w
wiitthh ooppttiioonn C
CO
OH
H22))::
C
CO
O content [ppm, mg/m³] 0...8,000ppm / 1ppm
±20ppm (0...200ppm), ±5% v. Mw. (201...2,000ppm),
±10% of reading (2001...8,000ppm) / t90 <40s
A
Am
mbbiieenntt C
Caarrbboonn m
moonnooxxiiddee,, vviiaa iinntteerrnnaall eelleeccttrroo--cchheem
miiccaall sseennssoorr::
A
Am
mbbiieenntt C
CO
O content
aam
mC
CO
O [ppm]
0...2,000ppm / 1ppm
±10ppm (0...100ppm), ±10% of reading (>100ppm) /
t90 <40s
TTeem
mppeerraattuurree,, vviiaa ttyyppee K
K tthheerrm
mooccoouuppllee ooff fflluuee ggaass pprroobbee ((N
NiiC
Crr--N
Nii))::
Flue gas temperature FFTT,
Flue gas dew point A
ATTP
P,
Ambient air A
ATT
-40...+600°C / 0.1°C,
±0.5°C (-40...100°C), ±0.5% of reading (>100°C),
-40...1,112°F / 0.1°F
±0.9°F (
-
40...212°F), ±0.5% of reading (>212°F) /
t98 <50s (TE 0.5mm); <100s (TE 1mm)
TTeem
mppeerraattuurree,, vviiaa ddiiffffeerreennttiiaall tteem
mppeerraattuurree sseett 00555544 11220088::
Flue gas socket T1 [°C, °F],
Sensor socket T2 [°C, °F] -40...+600°C / 0.1°C,
±0.5°C (-40...100°C), ±0.5% of reading (>100°C),
-40...1,112°F / 0.1°F
±0.9°F (
-
40...212°F), ±0.5% of reading (>212°F) /
t98 <50s (TE 0.5mm); <100s (TE 1mm)
P
Prreessssuurree,, vviiaa iinntteerrnnaall ddiiffffeerreennttiiaall pprreessssuurree sseennssoorr::
Flue draught D
Drrgghhtt [mbar,
hPa, inW, in Hg]
-
40...40hPa / 0.01hPa
±0.02hPa (-0.50...0.60hPa), ±0.03hPa
(0.61...3hPa), ±1.5% of reading (>3hPa) / -
Flue draught D
Drrgghhtt, with option
of precision draught [Pa]
-
100...100Pa / 0.1Pa
±3Pa / -
P
Prreessssuurree aanndd ttiigghhttnneessss tteesstt vviiaa iinntteerrnnaall ddiiffffeerreennttiiaall pprreessssuurree sseennssoorr,, w
wiitthh ddiiffffeerreennttiiaall pprreessssuurree sseett
00555544 11220033::
Differential pressure
∆∆
pp [hPa] -200...200hPa / 0.1hPa
±0.5hPa (0.0...50.0hPa)
(with option precision
±1% of reading (50.1...100.0hPa)
difference pressure: 0.01hPa) ±1.5% of reading (100.1...200.0hPa)
E
Effffiicciieennccyy,, ccaallccuullaatteedd::
Efficiency
ηη
2
[%],
Efficiency
ηη
+
+
3)
[%]
0...120% / 0.1%
±0.2% / -