Service recommendations

23

d

e

en

fr

es

it

p

t

sv

nl

??

??

Tips and assistance

22

Routine servicing, calibration and gas sensor replacement.

UK specific guideline BS7967-2 provides information on the use of electronic
portable gas analysers and their use including calibration requirements.
BS7967-2 (4.1) states that “Analysers should be treated with care and used
and maintained in accordance with the manufacturer’s instructions.” Testo Ltd
recommends that our Flue Gas Analysers are serviced and calibrated at least
once annually. 

Routine servicing and calibration is important to ensure that the instrument is
maintained properly and will continue to give reliable accurate readings. It is
also a requirement that any Flue Gas Analyser has current proof of calibration.
Proof of calibration is achieved by having an in date calibration certificate and
valid calibration label attached to the instrument. 

Testo Flue Gas Analysers use Electro/Chemical sensor technology to measure
gas concentrations and these sensors deplete not only with use but also with
time. Testo Ltd recommends that both the O

2

and CO sensors are replaced at

least once every 3 years by the Manufacturer or a Testo approved service
agent. The life and reliability of these sensors cannot be guaranteed beyond
this period and by having your instrument serviced and calibrated annually will
ensure that the sensors are checked and that they are measuring correctly.  

Testo Ltd will not, under any circumstances, be held liable for the performance
of any Testo gas analyser where the cell replacement, servicing or calibration
has been carried out by any organisation other than by Testo or a Testo
approved service agent. 

Warranties may be affected.

Tips and assistance

Questions and answers

M

Meeaassuurriinngg  iinnssttrruum

meenntt  ssw

wiittcchheess  iittsseellff  ooffff  oorr  w

wiillll  nnoott  ssw

wiittcchh  oonn??

· Rechargeable battery is low: charge battery or connect mains unit.
· On 327-2 models condensate trap is full, red lights indicate.

B

Baatttteerryy  ccaappaacciittyy  sseeeem

mss  ttoo  bbee  ffaauullttyy??

· Rechargeable battery was repeatedly not fully discharged/charged: discharge rechargeable

battery (until measuring instrument switches itself off) and then charge fully.

--------  aappppeeaarrss  iinnsstteeaadd  ooff  aa  rreeaaddiinngg??

· Sensor/probe is not plugged in: connect the sensor/probe
· Sensor/probe or measuring cell faulty: check sensor/probe or measuring cell.

M

Meessssaaggee::  P

Puum

mpp  fflloow

w  rraattee  ttoooo  hhiigghh??

· Gas outlet is blocked: make sure that the gas outlet is clear.

M

Meessssaaggee::  E

ER

RR

RO

OR

R  +

+  ttw

woo--ddiiggiitt  nnoo..  aanndd  sseerrvviiccee??

Device error: switch off the instrument and contact your dealer or Testo Customer Service.

If the above information does not solve the problem, please contact your dealer
or Testo Customer Service. For contact data, see back of this document or
web page www.testo.com/service-contact

Содержание 0563320370

Страница 1: ...Instruction manual en testo 327 Flue gas analyser country specific version GB...

Страница 2: ...he measuring instrument together with solvents and do not use any desiccants Only operate the measuring instrument properly for its intended purpose and within the parameters specified in the technica...

Страница 3: ...o c ch he em miic ca all s se en ns so or r n no ot t 3 32 27 7 C CO O O O2 2 content O O2 2a aiir r supply Reference value O O2 2 r re ef f 0 21 0 1 0 2 t90 40s C Ca ar rb bo on n m mo on no ox xiid...

Страница 4: ...urce Li ion rechargeable battery 3 7 V 1 4 Ah 0515 0114 3 7 V 2 4 Ah 0515 0100 mains unit 6 3V 1 2A Battery life measuring gas pump on display light off Approx 4 h 0515 0114 approx 10h 0515 0100 Batte...

Страница 5: ...blle e b ba at tt te er ry y iin n c ch ha ar rg ge er r 0 05 55 54 4 1 10 08 87 7 a ac cc ce es ss so or ry y Refer to the documentation that comes with the charger de en fr es it pt sv nl Product de...

Страница 6: ...ut applying the changes E Ex xa am mp plle e C Ch ha an ng ge e d diis sp plla ay y p po os siit tiio on n The position number to be changed has been selected 1 Press several times until the required...

Страница 7: ...and re start unit C Co on nn ne ec ct tiin ng g t th he e p pr ro ob be es s Insert the connector into the flue gas socket and lock by turning it clockwise gently bayonet lock There must be no more th...

Страница 8: ...iit th ho ou ut t o op pe er ra at tiin ng g p pr re es ss su ur re e is displayed After zeroing the message A At tt ta ac ch h t to o t th he e s sy ys st te em m a an nd d s st ta ar rt t is display...

Страница 9: ...e g ga as s measurement menu using the O OK K function key The values entered are not shown in the central Flue gas measurement menu However they can be printed out together with the readings from ot...

Страница 10: ...cations that have already been assigned are indicated by the display of the date time of saving 1 1 Press 2 2 Select M Me em mo or ry y using The memory capacity and available memory locations are dis...

Страница 11: ...r r Check the particle filter of the flue gas probe for contamination at regular intervals check visually by looking through the window of the filter chamber Replace the filter if there are signs of c...

Страница 12: ...ng correctly Testo Ltd will not under any circumstances be held liable for the performance of any Testo gas analyser where the cell replacement servicing or calibration has been carried out by any org...

Страница 13: ...en ns so or rs s Compact flue gas probe 180 mm 6 mm TC 1 mm incl cone up to 500 C 932 F 0600 9740 Compact flue gas probe 300 mm 6 mm TC 1 mm incl cone up to 500 C 932 F 0600 9741 Thermocouple for comp...

Страница 14: ...CO O x x CO Measured carbon monoxide value Calculated air ratio F Fllu ue e g ga as s d de ew w p po oiin nt t t te em mp pe er ra at tu ur re e F FT TP P FH20 Flue gas specific water vapour level in...

Страница 15: ...Notes 29 de en fr es it pt sv nl Notes 28...

Страница 16: ...0977 3270 05 en V01 09 en_GB...

Отзывы: