First steps
10
33
Select the required function using / and confirm entry with the O
OK
K
function key.
FFuunnccttiioonnss
11.. D
Diissppll.. sseeqq: selects parameters and units of measurement and assigns a position number
for the
display/protocol printouts.
22.. D
Daattee//TTiim
mee: sets the date and time
33.. LLaanngguuaaggee: sets the language.
44.. P
Prriinntteerr: sets the printer to be used.
- The selected function is opened and the position number
(D
Diissppll.. sseeqq function only) or parameter which can be set flashes.
55.. B
Blluueettooootthh (327-2 with option Bluetooth only): activate/ deactivate interface.
44
Set the position number (D
Diissppll.. sseeqq function only)/parameter:
For D
Diissppll.. sseeqq function only: Select the position number to be changed
using / and confirm with the C
Chhaannggee function key.
Alternatively: Delete the position number with D
Deell.. and insert a new
position number using IInnss...
D
Diissppll.. sseeqq function: Only parameters and units of measurement which are
assigned to a position number appear in the display and on printouts.
A maximum of 20 position numbers can be activated.
K
Keeyy ffuunnccttiioonnss
·
: Change parameters.
·
For D
Daattee// TTiim
mee function: changes between hours, minutes, day, month and year.
·
For D
Diissppll.. sseeqq function: changes between parameter and unit of measurement (only
available if there are several units of measurement for the selected parameter).
· O
OK
K for D
Diissppll.. sseeqq function and a flashing position number: confirms setting and moves to
next display position.
· O
OK
K for D
Diissppll.. sseeqq and FFiinniisshh flashing: confirms settings and leaves the function.
· O
OK
K for D
Daattee// TTiim
mee, LLaanngguuaaggee, and P
Prriinntteerr: functions: confirms setting and leaves the
function.
· eesscc: leaves the parameter or function without applying the changes.
E
Exxaam
mppllee:: ""C
Chhaannggee ddiissppllaayy ppoossiittiioonn""
The position number to be changed has been selected.
1 Press
/
several times until the required parameter flashes.
2 Press the or
function key to go to the menu for selecting the unit of measurement.
3 Press
/
several times until the required unit of measurement flashes.
4 Press the O
OK
K function key to confirm the setting and move to the next display position.
5 At the end of performing the settings: Press
/
several times until FFiinniisshh flashes
(appears after the last position number) and confirm entry with O
OK
K.
Operation with the mains unit
If the mains unit is connected, the measuring instrument is automatically
powered from the mains unit. It is not possible to charge the rechargeable
battery in the measuring instrument during operation.
11
Connect the plug of the mains unit to the mains unit socket on the
measuring instrument.
22
Connect the mains plug of the mains unit to a mains socket.
- The measuring instrument is powered via the mains unit.
- If the instrument is switched off and a rechargeable battery is inserted,
the charging process will start automatically. Switching the measuring
instrument on has the effect of stopping rechargeable battery charging
and the measuring instrument is then powered via the mains unit.
Switching on/off
S
Sw
wiittcchhiinngg tthhee iinnssttrruum
meenntt oonn::
Press .
- Initialisation phase:
· All display segments are lit (length of time: 3 s).
· Serial number, firmware version, instrument designation, date, time
and country-specific version of instrument are displayed (length of
time: 5 s).
- The M
Meeaassuurree flue gas option is displayed.
S
Sw
wiittcchhiinngg tthhee iinnssttrruum
meenntt ooffff::
Press .
- sometimes: The pump starts and the measuring cells are rinsed until the
switch-off thresholds (O
2
>20 %, other parameters 50 ppm) are reached.
Rinsing lasts no more than 2 minutes.
Performing instrument settings
P
Peerrffoorrm
miinngg sseettttiinnggss::
11
Press .
22
Select S
Seettttiinnggss using / and confirm entry with the O
OK
K function key.