Maintaining the product

19

d

e

en

fr

es

it

p

t

sv

nl

??

??

Checking the instrument

18

Maintaining the product

Condensate trap

The fill level of the condensate trap can be read from the markings on the
condensate trap.

E

Em

mppttyyiinngg  tthhee  ccoonnddeennssaattee  ttrraapp

Flue Gas condensate consists of a weak mix of acids. Avoid contact with the
skin. Make sure that the condensate does not run over the housing.

Caution! Failure to empty the condensate trap when the fill level is
reached results in condensate entering the sensors and pump. This
voids the warranty and results in costly repairs and replacements.

Do not empty the condensate trap while the flue gas pump is in
operation!

11

Keep measuring instrument in an upright
position (condensate outlet pointing upwards).

22

Open the condensate outlet on the
condensate trap: pull out approx. 7 mm to
the stop.

33

Let the condensate run out into a sink.

44

Mop up any remaining drops on the
condensate outlet using a cloth.

55

Close the condensate outlet.

The condensate outlet must be completely

closed, otherwise measuring errors could occur if external air gets in. It is
advisable to empty water trap at regular intervals, (preferably daily).

P

Prriinnttiinngg//ddeelleettiinngg//ddiissppllaayyiinngg  m

meeaassuurreem

meenntt  ddaattaa  ((332277--22  oonnllyy))::

There are 20 memory locations (P

Poossiittiioonn  11 to P

Poossiittiioonn  2200) to which one

measurement data record can be saved per location. Memory locations 
that have already been assigned are indicated by the display of the 
date/time of saving.

11

Press  . 

22

Select M

Meem

moorryy using  /  .

- The memory capacity and available memory locations are displayed.

To print the memory: press the P

Prriinntt function key.

To delete the whole memory: press the D

Deell function key and confirm entry

with the Y

Yeess function key.

33

Press O

OK

K.

44

Select memory location using  /  .

To display the measurement data record: press the V

Vaalluuee function key.

To print the measurement data record: press the P

Prriinntt function key.

To delete the measurement data record: press the D

Deell function.

TTrraannssffeerr  ddaattaa  ttoo  aa  P

Poocckkeett  P

PC

C  ((332277--22  oonnllyy))::

Data can be transferred to a Pocket PC via infrared or Bluetooth.
You must also refer to the documentation that comes with the software. 

Checking the instrument

P

Peerrffoorrm

miinngg  aann  iinnssttrruum

meenntt  ddiiaaggnnoossiiss::

11

Press  . 

22

Select D

Diiaaggnnoossiiss using  /  and confirm entry with the O

OK

K function key.

33

Select the required function using  /  and confirm entry with the O

OK

K

function key.

FFuunnccttiioonnss

11.. IInnffoo: displays instrument information: serial number, instrument temperature, operating

hours, qA version, last service

22.. E

Errrroorr: displays list of errors.

33.. R

Reecchh..  bbaatttt: displays the battery capacity.

44.. S

Seennss..  D

Diiaagg (327-2 only): performs sensor diagnosis.

K

Keeyy  ffuunnccttiioonnss  ffoorr  S

Seennss..  D

Diiaagg  ffuunnccttiioonn  ((332277--22  oonnllyy))

·

: select sensor.

· R

Reeaadd: performs sensor diagnosis and displays results of diagnosis.

Содержание 0563320370

Страница 1: ...Instruction manual en testo 327 Flue gas analyser country specific version GB...

Страница 2: ...he measuring instrument together with solvents and do not use any desiccants Only operate the measuring instrument properly for its intended purpose and within the parameters specified in the technica...

Страница 3: ...o c ch he em miic ca all s se en ns so or r n no ot t 3 32 27 7 C CO O O O2 2 content O O2 2a aiir r supply Reference value O O2 2 r re ef f 0 21 0 1 0 2 t90 40s C Ca ar rb bo on n m mo on no ox xiid...

Страница 4: ...urce Li ion rechargeable battery 3 7 V 1 4 Ah 0515 0114 3 7 V 2 4 Ah 0515 0100 mains unit 6 3V 1 2A Battery life measuring gas pump on display light off Approx 4 h 0515 0114 approx 10h 0515 0100 Batte...

Страница 5: ...blle e b ba at tt te er ry y iin n c ch ha ar rg ge er r 0 05 55 54 4 1 10 08 87 7 a ac cc ce es ss so or ry y Refer to the documentation that comes with the charger de en fr es it pt sv nl Product de...

Страница 6: ...ut applying the changes E Ex xa am mp plle e C Ch ha an ng ge e d diis sp plla ay y p po os siit tiio on n The position number to be changed has been selected 1 Press several times until the required...

Страница 7: ...and re start unit C Co on nn ne ec ct tiin ng g t th he e p pr ro ob be es s Insert the connector into the flue gas socket and lock by turning it clockwise gently bayonet lock There must be no more th...

Страница 8: ...iit th ho ou ut t o op pe er ra at tiin ng g p pr re es ss su ur re e is displayed After zeroing the message A At tt ta ac ch h t to o t th he e s sy ys st te em m a an nd d s st ta ar rt t is display...

Страница 9: ...e g ga as s measurement menu using the O OK K function key The values entered are not shown in the central Flue gas measurement menu However they can be printed out together with the readings from ot...

Страница 10: ...cations that have already been assigned are indicated by the display of the date time of saving 1 1 Press 2 2 Select M Me em mo or ry y using The memory capacity and available memory locations are dis...

Страница 11: ...r r Check the particle filter of the flue gas probe for contamination at regular intervals check visually by looking through the window of the filter chamber Replace the filter if there are signs of c...

Страница 12: ...ng correctly Testo Ltd will not under any circumstances be held liable for the performance of any Testo gas analyser where the cell replacement servicing or calibration has been carried out by any org...

Страница 13: ...en ns so or rs s Compact flue gas probe 180 mm 6 mm TC 1 mm incl cone up to 500 C 932 F 0600 9740 Compact flue gas probe 300 mm 6 mm TC 1 mm incl cone up to 500 C 932 F 0600 9741 Thermocouple for comp...

Страница 14: ...CO O x x CO Measured carbon monoxide value Calculated air ratio F Fllu ue e g ga as s d de ew w p po oiin nt t t te em mp pe er ra at tu ur re e F FT TP P FH20 Flue gas specific water vapour level in...

Страница 15: ...Notes 29 de en fr es it pt sv nl Notes 28...

Страница 16: ...0977 3270 05 en V01 09 en_GB...

Отзывы: