Multi Station
Vicat Test Apparatus
51
6.1 ASTM-Devices
6.1.1 ASTM-Device
“1.0366.ASTM “
Figure 58: ASTM device
“1.0366.ASTM”
Note
To achieve faster desired humidity within the device switch off cooling while pump is
still working. Please press the indicated button 3 seconds. (“OFF”)
For the highest possible air humidity, it is better if the ambient temperature of the
room is lower than that set in the device (the device heats up).
Each time the appliance cools, the humidity drops slightly.
Note
If the humidity sensor shows 0%, it is either wet or dirty. If the humidity sensor is wet,
it must be carefully blown dry. If the sensor is dirty (lime, dirt), it must be carefully
cleaned with a soft brush and a non-aggressive cleaner (e.g. brake cleaner).
(The surface must look crystal clear).
foam
Humidity sensor
Venting valve with
tube
Connect
ASTM-Input
Temperature sensor
This device hasn´t got a fil-
ter