Multi Station
Vicat Test Apparatus
38
Figure 40: window for determining the zero levels of the samples
Press
button “1”
to start determining the zero levels of all samples.
The display changes and shows the progress of the zero levels determination (
Once completed, all zero levels will be saved to the EEPROM and will be immediately available for the
subsequent tests. The display shows the above window again (Fehler! Verweisquelle konnte nicht gefun-
den werden.).
Pressing
button “2”
display remains unchanged because there is no follow-up window.
Press
button "C"
to exit the selection window.
The needle moves into so-called parking position and the apparatus reports its readiness (
).
Figure 41: progress of the zero level measurement
Note
The glass plates must be on the testing stations to determine the zero levels!
The zero level is calculated incorrectly, if the rubber buffer is set so that the needle does
not reach the glass plate.
Note
Measurement deviation within one level is ±0.1mm (z-axis).
Measurement repeatability at the same point is ±0.05mm (z-axis).
Determine zero lev-
els of samples?
yes: 1 back: 3
no: 2 cancel C
Determine zero level
Sample: 1
Measuring point: 3