Multi Station
Vicat Test Apparatus
9
1.7 Connection to the power supply
Danger
In accordance with the pertinent standards, the connection socket should be con-
nected to the grounding system.
Also, check the device rating plate to make sure that the ratings of the building power
supply conform to the requirements for voltage, wattage, amperage, and frequency
of the equipment.
The technical characteristics of this safety device must also satisfy the standards that
apply in the country in which the device is installed.
Caution
The manufacturer of the equipment cannot be held liable for any damages that result
because the information here is not observed.
Electrical tolerances:
Actual voltage:
10% of the rated voltage
Frequency: 50 Hertz
1% of the rated frequency, continually;
2% of the rated frequency, on a
short-term basis
The manufacturer shall not be liable for damages to persons or properties that arise because the
above instructions have not been considered.
To protect the electronic components from short circuit, two troughs fuses (250VAC/ 6.3A) are build-in,
easily accessible to the user.
Figure 1: Fuses
Fuse
T 250VAC/6,3A
4, 6 and 8 stations:
12 stations
Fuse
T 250VAC/6,3A