Multi Station
Vicat Test Apparatus
16
Settings that can be made on the device:
Button: UP
By pushing this button, you increase the parameter or the parameter
value.
Button: DOWN
By pushing this button, you decrease the parameter or the parameter
value. In case of an alarm, you can also shut off the acoustic alert
(buzzer) by pressing this button.
Not active
Button: SET
Pressing this button displays the set point value for heating
(Default
value: 19.7 °C).
This button is also used to set the parameters.
Standby
Standby circuitry
The controller has three (3) operation levels:
Operator level no. 1:
Parameterization of the set point value:
You can directly select the set point value for heating by pressing the SET button (regardless of the standby
mode). If you keep pushing SET key and press UP and DOWN buttons, you can increase or decrease the
set point values. The newly set values will then apply for the control functions.
Operator level no. 2 (P=parameters):
If you press both the UP and DOWN button at the same time and hold them pressed for at least 4 seconds,
you will get into a parameter list for the control parameters (these parameters begin with P1). You can use
the UP button to move the list upward, and the DOWN button to scroll downward in this list. When you press
the SET button, this will display the value of a single parameter. If you then press the UP or the DOWN
button, while still holding down the SET button, then you can increase or decrease the value of the respective
parameter. When you release all buttons, the new value will be saved. If you take no action or press any
button for more than 60 seconds, the system will jump back to the first display.
The following is a brief description of some of the parameters:
Parameter
Description of the function
Setting range
P0
Actual value
—
P1
Desired value (set point) for cooling
–99.9 ... +99.9 K (20.0°C)
P2
Hysteresis control-contact 1
0.1 ... 99.0 K ( 0.1 °C)
P3
Hysteresis control-contact 2
0.1 ... 99.0 K ( 0.1 °C)
P4
Lower limit for set point value
–99°C ... P5
P5
Upper limit for set point value
P4 ... 999°C
P6
Correction of actual value
–20.0 ... +20.0 K
P19
Locking the keyboard buttons
0 = unlocked; 1 = locked
P30
Lower limit value for alarm
–99 ... 999°C/K
P31
Upper limit value for alarm
–99 ... 999°C/K
P32
Hysteresis alarm, one-sided
0.1 ... 99.9 K